Oxygen signal

Description

The oxygen signal of the O2k is transmitted from the electrochemical polarographic oxygen sensor (OroboPOS) for each of the two chambers to DatLab. The primary signal is a current mAmp which is converted into a voltage [V], and calibrated in SI units for amount of substrance concentration [ยตmol.dm-3 or ยตM].


Reference: MiPNet19.18D O2k-Calibration, MiPNet06.03 POS-Calibration-SOP



MitoPedia methods: Respirometry, "DatLab" is not in the list (Respirometry, Fluorometry, Spectrophotometry) of allowed values for the "MitoPedia method" property. DatLab 

Template:MitoPedia SUIT



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Stability of the oxygen signal

The stability of the oxygen signal of the O2k is evaluated by the (negative) uncorrected slope over time. After smoothing, the slope of a stable sensor should be less than +- 1 pmol.s-1.ml-1 in a 2-ml chamber.

  • Instability of the oxygen signal
  • There is drift of the oxygen signal over short or long periods of time.
  • The signal at air saturation is abnormally high for a given gain setting.
  • The zero current is high (>2.5%) and may show drift.
Follow the standard instructions for oxygen sensor service.