|abstract='''Gnaiger E (2016) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(15):1-8.''' Β» [http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions] | |abstract='''Gnaiger E (2016) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(15):1-8.''' Β» [http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions] | ||
'''[[O2k-SOP |O2k-Protocols SOP]]''': [[High-resolution respirometry]] (HRR; OROBOROS Oxygraph-2k, O2k) critically depends on maintenance ([[MiPNet19.18B POS- | '''[[O2k-SOP |O2k-Protocols SOP]]''': [[High-resolution respirometry]] (HRR; OROBOROS Oxygraph-2k, O2k) critically depends on maintenance ([[MiPNet19.18B POS-service]]) and accurate calibration of the [[polarographic oxygen sensor]]s ([[OroboPOS]], POS). Standard operating procedures (O2k-SOP) are described as part of the [[MitoFit Quality Control System]]. (1) Clean and ready-to-use O2k-Chambers are a basic requirement ([[MiPNet19.03]]). (2) Quality control for evaluation of proper POS function (SOP: O2 sensor test). (3) Accurate POS calibration ([[MiPNet19.18D O2k-calibration]]). This is part 1 of O2k Quality Control. | ||
:Β» O2k-Protocols SOP: O2k Quality Control 2 Β»[[MiPNet14.06 InstrumentalO2Background]]Β« | :Β» O2k-Protocols SOP: O2k Quality Control 2 Β»[[MiPNet14.06 InstrumentalO2Background]]Β« |
Revision as of 12:20, 14 August 2016
MiPNet06.03 POS-calibration-SOP
O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. |
OROBOROS (2016-02-09) Mitochondr Physiol Network
Abstract: Gnaiger E (2016) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(15):1-8. Β» Versions
O2k-Protocols SOP: High-resolution respirometry (HRR; OROBOROS Oxygraph-2k, O2k) critically depends on maintenance (MiPNet19.18B POS-service) and accurate calibration of the polarographic oxygen sensors (OroboPOS, POS). Standard operating procedures (O2k-SOP) are described as part of the MitoFit Quality Control System. (1) Clean and ready-to-use O2k-Chambers are a basic requirement (MiPNet19.03). (2) Quality control for evaluation of proper POS function (SOP: O2 sensor test). (3) Accurate POS calibration (MiPNet19.18D O2k-calibration). This is part 1 of O2k Quality Control.
- Β» O2k-Protocols SOP: O2k Quality Control 2 Β»MiPNet14.06 InstrumentalO2BackgroundΒ«
- Β» O2k-Manual Β»MiPNet19.18D_O2k-calibrationΒ«
- Β» Product: OroboPOS, O2k, O2k-Catalogue
- Β» Template: O2-calibration.xlsx
β’ O2k-Network Lab: AT_Innsbruck_OROBOROS
Labels: MiParea: Instruments;methods
HRR: Oxygraph-2k, O2k-Protocol
O2k-SOP
Further details
Next step - O2k Quality Control 2: Β»MiPNet14.06 InstrumentalO2BackgroundΒ«
References
- Forstner H, Gnaiger E (1983) Calculation of equilibrium oxygen concentration. In: Polarographic Oxygen Sensors. Aquatic and Physiological Applications. Gnaiger E, Forstner H (eds), Springer, Berlin, Heidelberg, New York:321-33. - Β»Bioblast linkΒ«
- Gnaiger_2008_POS
- MiPNet19.18D_O2k-calibration