Template Select fluorescence sensor
From Bioblast
Select the Fluo-Sensors
- Switching between different excitation wavelengths and filters is achieved by simply exchanging the Fluo-Sensors. Two types of optical sensors are supplied with different LEDs for fluorescence excitation, and the effective spectra of the LEDs are modified by filters.
- Select the Fluo-Sensor and filter set from the table on application-specific settings (here we should link to the respective section which is also a template). Each Fluo-Sensor is delivered with a mounted filter set:
- Fluorescence-Sensor Green and Smart Fluo-Sensor Green: 525 nm (dominant wavelength). The installed Filter-Cap is ready for H2O2 measurement with Amplexยฎ UltraRed or for measurements of mt-membrane potential with TMRM.
- Fluorescence-Sensor Blue and Smart Fluo-Sensor Blue: 465 nm (dominant wavelength). The installed Filter-Cap is ready for measurements of mt-membrane potential with safranin or Rhodamine 123. A different filter is used for measurement with Magnesium Greenโข or Calcium Greenโข.