Difference between revisions of "Template:Fluorescence dyes and settings"
From Bioblast
Line 3: | Line 3: | ||
! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage) !! Gain !! Comment | ! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage) !! Gain !! Comment | ||
|- | |- | ||
| [[Amplex UltraRed| | | [[Amplex UltraRed|Amplex<sup>TM</sup> UltraRed]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type II]] || 100 - 500 || 1000 || at c(AmR) = 10 µM | ||
|- | |- | ||
| [[TMRM]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type III]] || 200 - 500|| 1000 || at c(TMRM) = 2 µM | | [[TMRM]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type III]] || 200 - 500|| 1000 || at c(TMRM) = 2 µM |
Revision as of 15:25, 18 November 2021
Application | Sensor | Filter set | O2k output | Light intensity (Amp polarization voltage) | Gain | Comment |
---|---|---|---|---|---|---|
AmplexTM UltraRed | Fluorescence-Sensor Green | AmR | Type II | 100 - 500 | 1000 | at c(AmR) = 10 µM |
TMRM | Fluorescence-Sensor Green | AmR | Type III | 200 - 500 | 1000 | at c(TMRM) = 2 µM |
Safranin | Fluorescence-Sensor Blue | Saf | Type III | 100 - 200 | 1000 | at c(safranin) = 2 µM |
Magnesium Green™ | Fluorescence-Sensor Blue | MgG / CaG | Type I and II | 100 - 500 | 1000 | at c(Mg Green) = 2 µM |
Calcium Green™ | Fluorescence-Sensor Blue | MgG / CaG | Type I | 300 - 500 | 1000 | at c(Ca Green) = 2 µM |
In the O2k-LED2-Module (Series D-G) polarization voltage is used, while in the O2k-Fluo Smart Module (Series H-I) fluorescence intensity. To use the same setting in the O2k-Fluo LED2-Module as in the O2k-Fluo Smart-Module use the following Excel sheet:
Download the EXCEL calculation-sheet to adjust fluorescence intensity |