| == High zero current ==
| |
| :::: In order to localize the problem, the following test should be performed:
| |
|
| |
| :::[[File:High zero current test 2.png|1000px]]
| |
| ::::*Note: Before sending us the O2k or any spare components, please contact us via our O2k-Open Support system: [[O2k-Open Support]]
| |
| :::::References:
| |
| ::::::[1] [https://wiki.oroboros.at/index.php/OroboPOS-Connector_MitoPedia#Cleaning_the_electrical_connections Cleaning the electrical connections]
| |
| ::::::[2] [[MiPNet19.18B POS-service]]
| |
| ::::::[3] [[MiPNet06.03 POS-calibration-SOP]]
| |
| ::::::[4] [[MiPNet14.06 Instrumental O2 background]]
| |
|
| |
| ::: '''Example:'''
| |
| ::::[[File:High zero current good example.png|600px]]
| |
| ::::[[File:High zero current test bad exmaple.png|600px]]
| |
| ::::'''Figure 1.''' High zero current test.Β Upper panel: Zero current is in the range of Β±0.00099 V. Lower panel: Zero current is higher than 0.00099 V when the OroboPOS-Connector is connected to the O2k-Main unit and when the OroboPOS is connected to the OroboPOS-Connector, which means that both the OroboPOS-Connector and OroboPOS have electrical problems. First, the OroboPOS-Connector and the POS were disassembled and disconnected from the main unit. Second, the OroboPOS-Connector was connected to the O2k-Main unit without the OroboPOS. Third, the OroboPOS was connected to the OroboPOS-Connector without membranes. After each step, the O<sub>2</sub> raw signal [V] was recorded for 10 min. The marks were set to the O<sub>2</sub> raw signal [V] to evaluate the zero current.
| |