== High zero current == | |||
:::: In order to localize the problem, the following test should be performed: | |||
:::[[File:High zero current test 2.png|1000px]] | |||
::::*Note: Before sending us the O2k or any spare components, please contact us via our O2k-Open Support system: [[O2k-Open Support]] | |||
:::::References: | |||
::::::[1] [https://wiki.oroboros.at/index.php/OroboPOS-Connector_MitoPedia#Cleaning_the_electrical_connections Cleaning the electrical connections] | |||
::::::[2] [[MiPNet19.18B POS-service]] | |||
::::::[3] [[MiPNet06.03 POS-calibration-SOP]] | |||
::::::[4] [[MiPNet14.06 Instrumental O2 background]] | |||
::: '''Example:''' | |||
::::[[File:High zero current good example.png|600px]] | |||
::::[[File:High zero current test bad exmaple.png|600px]] | |||
::::'''Figure 1.''' High zero current test.ย Upper panel: Zero current is in the range of ยฑ0.00099 V. Lower panel: Zero current is higher than 0.00099 V when the OroboPOS-Connector is connected to the O2k-Main unit and when the OroboPOS is connected to the OroboPOS-Connector, which means that both the OroboPOS-Connector and OroboPOS have electrical problems. First, the OroboPOS-Connector and the POS were disassembled and disconnected from the main unit. Second, the OroboPOS-Connector was connected to the O2k-Main unit without the OroboPOS. Third, the OroboPOS was connected to the OroboPOS-Connector without membranes. After each step, the O<sub>2</sub> raw signal [V] was recorded for 10 min. The marks were set to the O<sub>2</sub> raw signal [V] to evaluate the zero current. |
Revision as of 08:27, 7 October 2020
Popular Bioblast page
- Zero calibration has been accessed more than
- 5,000 times (2019-12-16)
- Zero calibration has been accessed more than
High zero current
- In order to localize the problem, the following test should be performed:
-
- Note: Before sending us the O2k or any spare components, please contact us via our O2k-Open Support system: O2k-Open Support
-
- Example:
- Figure 1. High zero current test. Upper panel: Zero current is in the range of ยฑ0.00099 V. Lower panel: Zero current is higher than 0.00099 V when the OroboPOS-Connector is connected to the O2k-Main unit and when the OroboPOS is connected to the OroboPOS-Connector, which means that both the OroboPOS-Connector and OroboPOS have electrical problems. First, the OroboPOS-Connector and the POS were disassembled and disconnected from the main unit. Second, the OroboPOS-Connector was connected to the O2k-Main unit without the OroboPOS. Third, the OroboPOS was connected to the OroboPOS-Connector without membranes. After each step, the O2 raw signal [V] was recorded for 10 min. The marks were set to the O2 raw signal [V] to evaluate the zero current.
- Example: