Very low temperatures

Question: Is it possible to use the oxygraph with a chamber temperature of 0 Β°C?

Our system is specified down to 4 Β°C at 25 deg room temperature and and down to 2 Β°C at "lower ambient temperature", see specifications. Lower ambient temperature does not necessary mean a "cool room". For the 2 deg, an acclimatised room usually will be enough. While we can only guarantee the working of the instrument within these limits, it is possible to go to 0Β°C and has been done. In general, this may require a "cool room". To this purpose the main unit itself can be exposed to temperatures down to just above 0Β°C (all the electronic components are rated at least down to 0Β°C but obviously 0Β°C room temperature has to be avoided due to condensation concerns). Also of course for very low room temperatures above 0 deg humidity has to be controlled somehow. We did a quick test today and could reach 0Β°C chamber temperature quite easily at a room temperature of about 14 Β°C. To keep this temperature stable the cooling power of the Peltier element was already working at 95% of maximum, so some kind of cool room will probable be necessary