::::* Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa
::::* Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa


Β 
== Troubleshooting ==
:::: In rare cases, the barometric pressure is displayed as 0 in DatLab. In this case switch of the O2k and unplug it from the socket for at least 15 minutes. In most cases the problem is resolved after connecting the O2k to the socket and switching it on again. If the problem persists or happens again, please contact our [https://www.oroboros.at/index.php/o2k-technical-support/| technical support].


{{MitoPedia O2k and high-resolution respirometry
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry
}}
}}

Revision as of 12:30, 24 August 2023


                  


O2k-Open Support

O2k-Barometric Pressure Transducer



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

O2k-Barometric Pressure Transducer, A/D converter and digital output to DatLab for continuous recording of barometric pressure [kPa or mmHg], integrated into the air calibration of the OroboPOS (MiPNet06.03 POS-calibration-SOP). Integral component of the O2k-Main Unit. The warranty on the accuracy of the signal obtained from the O2k-Barometric Pressure Transducer expires within three years.


Reference: O2k-Core

SI unit: Pa = JΒ·m-3
1 atm = 760 mmHg = 101.325 kPa

Accuracy

Quick accuracy check for the barometric pressure transducer (Mario Faschung, 2013-10-21):
  • 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa
  • Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa

Troubleshooting

In rare cases, the barometric pressure is displayed as 0 in DatLab. In this case switch of the O2k and unplug it from the socket for at least 15 minutes. In most cases the problem is resolved after connecting the O2k to the socket and switching it on again. If the problem persists or happens again, please contact our technical support.


MitoPedia O2k and high-resolution respirometry: O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry