Difference between revisions of "O2k-Barometric Pressure Transducer"
From Bioblast
Beno Marija (talk | contribs) |
|||
Line 1: | Line 1: | ||
{{ | {{Technical support}} | ||
__TOC__ | |||
{{MitoPedia without banner | |||
|abbr= | |||
|description='''O2k-Barometric Pressure Transducer''', A/D converter and digital output to DatLab for continuous recording of [[barometric pressure]] [kPa or mmHg], integrated into the air calibration of the OroboPOS ([[MiPNet06.03 POS-calibration-SOP]]). Integral component of the [[O2k-Main Unit]]. The warranty on the accuracy of the signal obtained from the O2k-Barometric Pressure Transducer expires within three years. | |description='''O2k-Barometric Pressure Transducer''', A/D converter and digital output to DatLab for continuous recording of [[barometric pressure]] [kPa or mmHg], integrated into the air calibration of the OroboPOS ([[MiPNet06.03 POS-calibration-SOP]]). Integral component of the [[O2k-Main Unit]]. The warranty on the accuracy of the signal obtained from the O2k-Barometric Pressure Transducer expires within three years. | ||
|info=[[O2k-Core]] | |info=[[O2k-Core]] | ||
}} | }} | ||
Line 14: | Line 15: | ||
::::* 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa | ::::* 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa | ||
::::* Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa | ::::* Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa | ||
{{MitoPedia O2k and high-resolution respirometry | |||
|mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry | |||
}} |
Revision as of 12:25, 14 April 2021
O2k-Barometric Pressure Transducer |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
O2k-Barometric Pressure Transducer, A/D converter and digital output to DatLab for continuous recording of barometric pressure [kPa or mmHg], integrated into the air calibration of the OroboPOS (MiPNet06.03 POS-calibration-SOP). Integral component of the O2k-Main Unit. The warranty on the accuracy of the signal obtained from the O2k-Barometric Pressure Transducer expires within three years.
Reference: O2k-Core
- SI unit: Pa = JΒ·m-3
- 1 atm = 760 mmHg = 101.325 kPa
Accuracy
- Quick accuracy check for the barometric pressure transducer (Mario Faschung, 2013-10-21):
- 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa
- Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa