MiPNet19.18D O2k-Series G and DatLab 6: Calibration: Difference between revisions

From Bioblast
No edit summary
No edit summary
Line 1: Line 1:
{{Publication
{{Publication
|title=Gnaiger E (2007) Oxygen and pX (pH) Calibration by DatLab. Mitochondr Physiol Network 12.08.
|title=Gnaiger E (2013) Oxygen and pX (pH) Calibration by DatLab. Mitochondr Physiol Network 12.08.
|info='''Open Access - [[File:MiPNet12.08 O2k-Calibration.pdf]]''', [http://www.oroboros.at/index.php?O2k-o2calibrationΒ  MiPNet12.08 Open Access]
|info='''Bioblast [[File:MiPNet12.08 O2k-Calibration.pdf]]''', [http://www.oroboros.at/index.php?O2k-o2calibrationΒ  MiPNet12.08 Open Access]
|authors=OROBOROS
|authors=OROBOROS
|year=2007
|year=2013
|journal=Mitochondr Physiol Network
|journal=Mitochondr Physiol Network
|abstract=[[Accuracy|Accurate]] calibration of the oxygen sensor depends on (1)
|abstract=[[Accuracy|Accurate]] calibration of the oxygen sensor depends on (1)
Line 23: Line 23:
}}
}}
{{Labeling
{{Labeling
|area=Respiration, Instruments;methods
|area=Instruments;methods
|instruments=Oxygraph-2k, O2k-Manual, Protocol
|instruments=O2k-Manual, Protocol
|additional=O2k-SOP, DatLab
|additional=O2k-SOP, DatLab
|articletype=Protocol; Manual, MiPNet-online Publication
|articletype=Protocol; Manual, MiPNet-online Publication

Revision as of 01:20, 9 October 2013

Publications in the MiPMap
Gnaiger E (2013) Oxygen and pX (pH) Calibration by DatLab. Mitochondr Physiol Network 12.08.

Β» Bioblast File:MiPNet12.08 O2k-Calibration.pdf, MiPNet12.08 Open Access

OROBOROS (2013) Mitochondr Physiol Network

Abstract: Accurate calibration of the oxygen sensor depends on (1) calibration solutions prepared at known oxygen partial pressures, as achieved in the Oxygraph-2k (O2k) at defined temperature, continuously recorded total gas pressure (barometric pressure), and thermodynamic equilibrium between the gas and aqueous phase; (2) high stability of the signal of the polarographic oxygen sensor (POS), tested for sufficiently long periods of time; (3) linearity of signal output with oxygen pressure, achieved with the POS in the range between oxygen saturation and zero oxygen pressure; and (4) application of accurate oxygen solubility for aqueous solutions for the conversion of partial oxygen pressure into oxygen concentration (MiPNet06.03). The standard oxygen calibration procedure is described for O2k high-resolution respirometry with the automatic calibration routine of DatLab. β€’ Keywords: HRR, DatLab

β€’ O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Instruments;methods 





HRR: O2k-Manual, Protocol"Protocol" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property. 

O2k-SOP, DatLab 

Info

  • Version 6: 2012-10-24
  • Version 1: 2007
  • MiPNet08.07, MiPNet09.08 and MiPNet02.01 (DatLab 2) are integrated into MiPNet12.08.
Cookies help us deliver our services. By using our services, you agree to our use of cookies.