Difference between revisions of "MiPNet06.03 POS-calibration-SOP"
From Bioblast
(Undo revision 189025 by Passrugger Manuela (talk)) |
|||
Line 9: | Line 9: | ||
{{MiPNet pdf page linking to MitoPedia}} | {{MiPNet pdf page linking to MitoPedia}} | ||
|mipnetlab=AT_Innsbruck_Oroboros | |mipnetlab=AT_Innsbruck_Oroboros | ||
}} | |||
{{Labeling | |||
|area=Respiration, Instruments;methods | |||
|instruments=Oxygraph-2k, O2k-Manual | |||
|additional=DatLab, O2k-SOP | |||
}} | }} |
Revision as of 16:38, 7 January 2020
MiPNet06.03 POS-calibration-SOP
O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. |
ยป ยป Versions
Oroboros (2018-01-07) Mitochondr Physiol Network
Abstract: Gnaiger E (2018) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(17):1-20.
- ยป Current O2k-series: NextGen-O2k Series XB and O2k Series J
- ยป Current software versions DatLab 8.0: MitoPedia: DatLab
- Further details: ยป MitoPedia: O2k-Open Support
โข O2k-Network Lab: AT_Innsbruck_Oroboros
Labels: MiParea: Respiration, Instruments;methods
HRR: Oxygraph-2k, O2k-Manual
DatLab, O2k-SOP