MiPNet06.03 POS-calibration-SOP
O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. |
ยป ยป Versions
Oroboros (2018-01-07) Mitochondr Physiol Network
Abstract: Gnaiger E (2018) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(17):1-20.
- ยป Current O2k-series: NextGen-O2k Series XB and O2k Series J
- ยป Current software versions DatLab 8.0: MitoPedia: DatLab
- Further details: ยป MitoPedia: O2k-Open Support
โข O2k-Network Lab: AT_Innsbruck_Oroboros
Labels: MiParea: Instruments;methods
HRR: Oxygraph-2k, O2k-Manual
DatLab, O2k-SOP