high-resolution terminology - matching measurements at high-resolution


Layout for DatLab graphs

Description

A Layout in DatLab is a predefined selection of Plots to be displayed at specific scalings and in a specific order. Any defined graph layout can be selected from the Layout menu for defining initial settings for the plots [Ctrl+F6] and scaling [F6]. A layout can be modified and saved under a specific name. In typical standard DatLab graph layouts plots are shown for the left chamber in Graph 1 (top), and for the right chamber in Graph 2 (below). During data acquisition, a 30 min time range is frequently used and can be adjusted any time during the experiment [F6].


Reference: MiPNet19.18C DatLab-guide

Graph-menu.png
Layout-menu.png

MitoPedia O2k and high-resolution respirometry: DatLab 

Tabs

ยป Plots
ยป Scaling
ยป Info


Reference Layouts: O2k-FluoRespirometer

Version: DatLab 7

Summary

01 Calibration show Temp
02 Calibration - Background
03 Background high O2
----
04a Flux per volume
04b Flux per volume overlay
----
05a Specific flux
05b Specific flux overlay
06a Specific flux high O2
06b Specific flux high O2 overlay
----
07a Flux Control Ratios
07b Flux Control Ratios overlay

01 Calibration show Temp

General information
Layouts 1 ff. are considered for O2k-Core applications.
Layouts Amp* or pX* are selected for O2k-MultiSensor applications.
1 Calibration show temp shows O2 concentration and the negative slope in Graph 1 and 2, and temperature and Peltier power in Graph 3: This is the default layout used after switching on the O2k.
Graph 1 & 2: O2 concentration (blue lines, left Y-axis) and O2 slope neg. (uncorrected; red lines, right Y-axis) are displayed, in the top graph for the left chamber (A), and below for the right chamber (B).
Graph 3: The third graph (bottom) shows the block temperature on the left Y axis (red) and the Peltier power on the right Y axis (green). Only when both temperature and Peltier power are constant, the chambers have reached thermal equilibrium.
  1. During this phase a stirrer test can be performed [F9].
  2. The next step is to observe equilibration of the oxygen signal with a defined gas phase above the stirred aqueous phase ('open') chamber; usually with air as the first step to perform an oxygen sensor calibration.
ยป O2k-Manual: MiPNet19.18A_O2k-start


Reference Layouts: O2 & Amp

Version: DatLab 7
01 Amp Amperometric_Raw signal
02 Amp Amperometric_Calibrated


Reference Layouts: O2 & pX

Version: DatLab 7
01 Potentiometric
----
02a TPP_calibration
02b TPP_with_O2flux
02c TPP_calibrated_with_O2flux


Reference Layouts: Other

Version: DatLab 7
O2 raw signal